Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ottobock Modular Polycentric EBS Knee Joint 3R60 Recalled by Otto Bock Healthcare GmbH Due to Otto Bock HealthCare GmbH is recalling Modular Polycentric...

Date: January 4, 2016
Company: Otto Bock Healthcare GmbH
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Otto Bock Healthcare GmbH directly.

Affected Products

Ottobock Modular Polycentric EBS Knee Joint 3R60, 3R60=ST, 3R60=KD, 3R60=HD. Used for the prosthetic fitting of amputations of the lower. limb.

Quantity: 40 units

Why Was This Recalled?

Otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loctite was used during repair.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Otto Bock Healthcare GmbH

Otto Bock Healthcare GmbH has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report