Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Mobile Mini C-arm system Part # 1000-0001. Intended to provide Recalled by Orthoscan, Inc. Due to OrthoScan Inc. discovered during investigation of a non-standard...

Date: August 31, 2017
Company: Orthoscan, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Orthoscan, Inc. directly.

Affected Products

Mobile Mini C-arm system Part # 1000-0001. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.

Quantity: 2012 total

Why Was This Recalled?

OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Orthoscan, Inc.

Orthoscan, Inc. has 11 total recalls tracked by RecallDetector.

Related Recalls