Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Orthopedic Alliance Spine System (alias "Blue & Gold") Part Numbers: Recalled by Orthopedic Alliance LLC Due to Orthopedic Alliance is recalling Blue & Gold Implants...

Date: March 7, 2013
Company: Orthopedic Alliance LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Orthopedic Alliance LLC directly.

Affected Products

Orthopedic Alliance Spine System (alias "Blue & Gold") Part Numbers: BGC3010 Screw Cap BGR0040 Straight Rod 5.5 x 40mm BGR0050 Straight Rod 5.5 x 50mm BGR0060 Straight Rod 5.5 x 60mm BGR0070 Straight Rod 5.5 x 70mm BGR0080 Straight Rod 5.5 x 80mm BGR0090 Straight Rod 5.5 x 90mm BGR0100 Straight Rod 5.5 x 100mm BGR0110 Straight Rod 5.5 x 110mm BGR0120 Straight Rod 5.5 x 120mm BGR0130 Straight Rod 5.5 x 130mm BGR0140 Straight Rod 5.5 x 140mm BGR0150 Straight Rod 5.5 x 150mm BGR0200 Straight Rod 5.5 x 200mm BGR0300 Straight Rod 5.5 x 300mm BGR0400 Straight Rod 5.5 x 400mm BGR0500 Straight Rod 5.5 x 500mm BGS5535 5.5 x 35mm Polyaxial Screw BGS5540 5.5 x 40mm Polyaxial Screw BGS5545 5.5 x 45mm Polyaxial Screw BGS6035 6.0 x 35mm Polyaxial Screw BGS6040 6.0 x 40mm Polyaxial Screw BGS6045 6.0 x 45mm Polyaxial Screw BGS6050 6.0 x 50mm Polyaxial Screw BGS6535 6.5 x 35mm Polyaxial Screw BGS6540 6.5 x 40mm Polyaxial Screw BGS6545 6.5 x 45mm Polyaxial Screw BGS6550 6.5 x 50mm Polyaxial Screw BGS7035 7.0 x 35mm Polyaxial Screw BGS7040 7.0 x 40mm Polyaxial Screw BGS7045 7.0 x 45mm Polyaxial Screw BGS7050 7.0 x 50mm Polyaxial Screw BGS7055 7.0 x 55mm Polyaxial Screw BGS7535 7.5 x 35mm Polyaxial Screw BGS7540 7.5 x 40mm Polyaxial Screw BGS7545 7.5 x 45mm Polyaxial Screw A pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Quantity: 3,300 units

Why Was This Recalled?

Orthopedic Alliance is recalling Blue & Gold Implants and Instruments because the Blue & Gold system is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product peformance failures that could cause patient harm due to implant breakage, loosening, or inadequate sterilization.

Where Was This Sold?

Distributed in Texas.

About Orthopedic Alliance LLC

Orthopedic Alliance LLC has 10 total recalls tracked by RecallDetector.