Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

OrthoPediatrics PediLoc Locking Plate System Recalled by OrthoPediatrics Corp Due to The device is a Left 8 hole plate...

Date: June 14, 2012
Company: OrthoPediatrics Corp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact OrthoPediatrics Corp directly.

Affected Products

OrthoPediatrics PediLoc Locking Plate System, 4.5mm Contour Locking Compression Femur 8 Hole Plate Right. Used for pediatric patients as indicated for pelvic, small and long bone fractures.

Quantity: 15 devices

Why Was This Recalled?

The device is a Left 8 hole plate but is incorrectly labeled and etched as a Right device.

Where Was This Sold?

This product was distributed to 7 states: FL, GA, IN, MI, MO, OH, SC