Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Orthofix Pillar SA Ti Spacer System Intervertebral Body Fusion Spinal Device Recalled by Orthofix U.S. LLC Due to Mislabeling

Date: February 20, 2025
Company: Orthofix U.S. LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Orthofix U.S. LLC directly.

Affected Products

Orthofix Pillar SA Ti Spacer System Intervertebral Body Fusion Spinal Device, sterile, implant grade titanium alloy: (1) REF 82-3300SP, 33mm W x 24mm D x 10mm H, 7 degrees; (2) REF 82-3700SP, 37mm W x 28mm D x 10mm H, 7 degrees; and (3) REF 82-4000SP, 40mm W x 32mm D x 10mm H, 7 degrees.

Quantity: 20

Why Was This Recalled?

The product is mislabeled with the incorrect anterior height of 10mm, but the laser marking on the implant and corresponding trial both show the correct anterior height of 10.5mm.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Orthofix U.S. LLC

Orthofix U.S. LLC has 9 total recalls tracked by RecallDetector.

Related Recalls