Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack (Unique Recalled by Ortho-Clinical Diagnostics Due to When testing was conducted using certain lots, VITROS¿...

Date: February 18, 2016
Company: Ortho-Clinical Diagnostics
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

1) VITROS¿ Immunodiagnostic Products Total Beta-hCG II Reagent Pack (Unique Device Identifier No. 10758750002320, 20758750002327) and the 2)VITROS¿ Immunodiagnostic Products Total Beta-hCG II Calibrators (Unique Device Identifier No. 10758750002337, 20758750002334); For quantitative measurement of human chorionic gonadotropin (hCG) and its beta subunit in human serum and plasma.

Quantity: 14,780 pieces

Why Was This Recalled?

When testing was conducted using certain lots, VITROS¿ Systems generated results within the measuring range for samples known to not contain measurable hCG. The system reported results up to approximately 7.40 mIU/mL (IU/L) for patient samples that should have been less than the measuring range of the assay (<2.39 mIU/mL (IU/L)).

Where Was This Sold?

This product was distributed nationwide across all 50 states.