Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 Recalled by Ortho-Clinical Diagnostics Due to CRP samples diluted using VITROS Chemistry Products Specialty...

Date: October 26, 2016
Company: Ortho-Clinical Diagnostics
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740

Quantity: 124,063 units

Why Was This Recalled?

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics has 185 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report