Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

VITROS Chemistry Products Cl- Slides Recalled by Ortho-Clinical Diagnostics Due to The slides do not meet the maximum allowable...

Date: July 11, 2018
Company: Ortho-Clinical Diagnostics
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum and plasma using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Quantity: 128196

Why Was This Recalled?

The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics has 185 total recalls tracked by RecallDetector.