Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. Recalled by Ortho-Clinical Diagnostics Due to Potential to cause biased results in the upper...

Date: May 8, 2023
Company: Ortho-Clinical Diagnostics
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM. Product Code: 6801704

Quantity: 2400 units

Why Was This Recalled?

Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.