Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS 5 Recalled by Ortho-Clinical Diagnostics Due to software automatically switches the lot when a vancomycin...

Date: May 21, 2012
Company: Ortho-Clinical Diagnostics
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

VITROS 5,1 FS System Software For use in the in-vitro quantitative, semi quantitative, and qualitative measurement of variety of analytes of clinical interest.

Quantity: 2044

Why Was This Recalled?

software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.