Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

The VITROS¿ Chemistry Products Calibrator Kit 31 Recalled by Ortho-Clinical Diagnostics Due to There is an incorrect value (data/calibration mathematics) on...

Date: April 19, 2016
Company: Ortho-Clinical Diagnostics
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.

Affected Products

The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Quantity: 520 Units

Why Was This Recalled?

There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ortho-Clinical Diagnostics

Ortho-Clinical Diagnostics has 185 total recalls tracked by RecallDetector.