Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS XT 7600 Integrated System with Version 3.7.2 and below Recalled by Ortho-Clinical Diagnostics, Inc. Due to Performing a system shutdown on VITROS XT 3400...

Date: July 19, 2022
Company: Ortho-Clinical Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Quantity: 1148 units

Why Was This Recalled?

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

Where Was This Sold?

This product was distributed nationwide across all 50 states.