Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products Recalled by Ortho-Clinical Diagnostics, Inc. Due to VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results...

Date: February 12, 2021
Company: Ortho-Clinical Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

VITROS SARS-CoV-2 Antigen Reagent Pack - The VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Reagent Pack when used in combination with the VITROS Immunodiagnostic Products SARS-CoV-2 Antigen Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) specimens. For emergency authorization use only (EAU). Product Code: 6199941. OUS Product.

Quantity: 12177 units

Why Was This Recalled?

VITROS Immunodiagnostic Products SARS-CoV-2 Antigen False Reactive Results Associated When Using Remel M4RT Viral Transport Media

Where Was This Sold?

This product was distributed nationwide across all 50 states.