Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198 Recalled by Ortho-Clinical Diagnostics, Inc Due to Falsely high or delayed patient and QC results...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc directly.
Affected Products
VITROS Immunodiagnostic Products LH Reagent Pack Product Code: 1350198
Quantity: 1575 units total: 1183 US; 392 OUS
Why Was This Recalled?
Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ortho-Clinical Diagnostics, Inc
Ortho-Clinical Diagnostics, Inc has 3 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report