Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the Recalled by Ortho-Clinical Diagnostics, Inc. Due to Lower than expected iPTH patient and Quality Control...

Date: November 17, 2022
Company: Ortho-Clinical Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

VITROS Immunodiagnostic Products Intact PTH Reagent Pack- IVD For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) Catalog Number: 6802892

Quantity: 4,158 units

Why Was This Recalled?

Lower than expected iPTH patient and Quality Control (QC) results, reported negatively biased iPTH results- an average -12% shift in patient sample results

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ortho-Clinical Diagnostics, Inc.