Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures Recalled by Ortho-Clinical Diagnostics, Inc. Due to May Generate Negatively Biased Results When Testing Samples...

Date: October 11, 2024
Company: Ortho-Clinical Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740

Quantity: 7717 units

Why Was This Recalled?

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

Where Was This Sold?

This product was distributed nationwide across all 50 states.