Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate Recalled by Ortho-Clinical Diagnostics, Inc. Due to Kit assigned an extended expiration date of 24-Jul-2025...

Date: May 14, 2024
Company: Ortho-Clinical Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344

Quantity: 1718

Why Was This Recalled?

Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.

Where Was This Sold?

This product was distributed nationwide across all 50 states.