Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 Recalled by Ortho-Clinical Diagnostics, Inc. Due to A software defect affecting VITROS Systems running VITROS...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.
Affected Products
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Quantity: 15 US; 43 OUS
Why Was This Recalled?
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ortho-Clinical Diagnostics, Inc.
Ortho-Clinical Diagnostics, Inc. has 62 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report