Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ Recalled by Ortho-Clinical Diagnostics, Inc. Due to On-Analyzer Stability (OAS) time, supported for 6 weeks;...

Date: May 26, 2022
Company: Ortho-Clinical Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics, Inc. directly.

Affected Products

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems. Product Code: 6199960

Quantity: 20858

Why Was This Recalled?

On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

Where Was This Sold?

This product was distributed nationwide across all 50 states.