Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
enGen Laboratory Automation Systems (enGen) with TCAutomation -- For in Recalled by Ortho-Clinical Diagnostics Due to Software anomaly: potential for delay in the reporting...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Ortho-Clinical Diagnostics directly.
Affected Products
enGen Laboratory Automation Systems (enGen) with TCAutomation -- For in vitro quantitative measurement of a variety of analytes of clinical interest; for pre-analytical and post-analytical sample and data management. --- Background: The enGen Lab Automation System is a modification to VITROS Analyzers that includes the addition of an automated track sub-system (TCAutomation) for pre-analytical and post-analytical sample and data management. In addition to an Automated Track Sub-system, the enGen System includes one or more VITROS Chemistry, Immunodiagnostic and/or Integrated Systems (clinical analyzers) as well as a Data Management Subsystem (Middleware), and a Software Configuration (gsb) File. The enGen System is typically interfaced with a Laboratory Information System (LIS) via connectivity facilitated at the Data Management Subsystem level. The Bypass Module is an analyzer specific module that interfaces clinical analyzers to the enGen System and allows for aspirating patient samples directly from the clinical analyzers. A sample tube carrier sensor in the Bypass Module is used to detect sample tube carriers as they leave the metering location of the Bypass. If a tube carrier is not detected within a defined period of time (6.4 seconds) the next sample to be metered may not advance for processing.
Quantity: Domestic - 31 units; Foreign - 77 units
Why Was This Recalled?
Software anomaly: potential for delay in the reporting of patient sample test results when using enGen with TCAutomation Software v2.6 , v3.2 and v3.5 Configured with Bypass Modules for ADVIA Centaur, Abbott Architect, VITROS and/or Tosoh AIA Systems.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Ortho-Clinical Diagnostics
Ortho-Clinical Diagnostics has 185 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report