Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD Recalled by Ortho Solutions Inc Due to Ortho Solutions Inc is initiating a recall on...

Date: July 14, 2015
Company: Ortho Solutions Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ortho Solutions Inc directly.

Affected Products

Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Sterile Orthopaedic Bone Screw

Quantity: 875 units

Why Was This Recalled?

Ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence there is a risk that a screw may not lock onto the plate at the time of primary insertion into the plate and the bone. In the event of this occurring the non functioning screw would need to be replaced with an alternativ

Where Was This Sold?

This product was distributed to 9 states: CA, CO, GA, IL, LA, NY, NC, TX, WA

Affected (9 states)Not affected

About Ortho Solutions Inc

Ortho Solutions Inc has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report