Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Capnostream20 (INTL) N REFURBISHED Recalled by Oridion Medical 1987 Ltd. Due to The date/time, nurse call and alarm settings of...

Date: July 31, 2018
Company: Oridion Medical 1987 Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Oridion Medical 1987 Ltd. directly.

Affected Products

Capnostream20 (INTL) N REFURBISHED, CS78652

Quantity: 5

Why Was This Recalled?

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Oridion Medical 1987 Ltd.