Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Vitek 2 reagent cards: a) GN ID Recalled by Organon Teknika Inc Due to This notice has been initiated due to potential...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Organon Teknika Inc directly.
Affected Products
Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID, REF 21347, d) AST-GNOR, REF 22000, e) AST-GP67, REF 22226, f) AST-P612, REF 22359, g) AST-XN01, REF 410025, h) AST-N222, REF 413083, i) AST-N223, REF 413110, j) AST-N226, REF 413143, k) AST-N248, REF 413397, l) ASR-GN67, REF 413399, m) AST-GN69, REF 413400, n) AST-GN73, REF 413404, o) AST-GN79, REF 413436, p) AST-GN82, REF 413439, q) AST-N261, REF 413753, r) AST-XN06, REF 413944, s) AST-N275, REF 414313, t) AST-GP75, REF 415670, u) AST-N300, REF 416241, v) AST-N308, REF 413913, w) AST-N314, REF 417027, x) AST-N318, REF 417952, y) AST-P643, REF 418671, z) AST-YS08, REF 420739, aa) AST-N348, REF 420856, bb) AST-ST02, REF 420915, cc) AST-ST03, REF 421040, dd) AST-N355, REF 421351, ee) AST-P654, REF 421912, ff) AST-P655, REF 421913, gg) AST-N371, REF 422024, hh) AST-N372, REF 422241, ii) AST-N379, REF 423052,
Quantity: 129,321 cartons
Why Was This Recalled?
This notice has been initiated due to potential for compromised test card Top Seal Integrity which could: - yield false resistance for antibiotics on the AST panel - cause a false negative ESBL Test - result in a false positive urea (URE) reaction on ID cards
Where Was This Sold?
Worldwide distribution.
About Organon Teknika Inc
Organon Teknika Inc has 1 total recall tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report