Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System Recalled by Olympus Scientific Solutions Americas Due to Trigger lock option does not always prevent x-ray...

Date: June 22, 2016
Company: Olympus Scientific Solutions Americas
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Olympus Scientific Solutions Americas directly.

Affected Products

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Quantity: 2, 210 units distributed in the USA

Why Was This Recalled?

Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Olympus Scientific Solutions Americas

Olympus Scientific Solutions Americas has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report