Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for Recalled by Olympus Corporation of the Americas Due to Non-conformances in Veran devices (ig4 Image Guided System,...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Olympus Corporation of the Americas directly.
Affected Products
Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes INS-5700 SPiN EWC Biopsy Guide Kit
Quantity: 6484 units
Why Was This Recalled?
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Olympus Corporation of the Americas
Olympus Corporation of the Americas has 378 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report