Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Recalled by Olympus Corporation of the Americas Due to Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the...

Date: May 26, 2023
Company: Olympus Corporation of the Americas
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Olympus Corporation of the Americas directly.

Affected Products

Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube

Quantity: 686 units

Why Was This Recalled?

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Olympus Corporation of the Americas

Olympus Corporation of the Americas has 378 total recalls tracked by RecallDetector.