Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

OES 4000 Hysteroscope Recalled by Olympus Corporation of the Americas Due to The required leakage test was not performed following...

Date: April 18, 2024
Company: Olympus Corporation of the Americas
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Olympus Corporation of the Americas directly.

Affected Products

OES 4000 Hysteroscope, Model/Catalog Number: A4674A

Quantity: 1 unit

Why Was This Recalled?

The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Olympus Corporation of the Americas

Olympus Corporation of the Americas has 378 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report