Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper Recalled by Olympus Corporation of the Americas Due to Part (A-rubber) intended for a different model of...

Date: April 22, 2022
Company: Olympus Corporation of the Americas
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Olympus Corporation of the Americas directly.

Affected Products

EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190

Quantity: 8 units

Why Was This Recalled?

Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Olympus Corporation of the Americas

Olympus Corporation of the Americas has 378 total recalls tracked by RecallDetector.

Related Recalls