Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Giraffe Exam Light The Exam Light illuminates body surfaces and Recalled by Ohmeda Medical Due to Certain power cords could overheat. The affected power...

Date: November 18, 2016
Company: Ohmeda Medical
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ohmeda Medical directly.

Affected Products

Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.

Quantity: 5,993 (5,080 US; 913 OUS)

Why Was This Recalled?

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ohmeda Medical

Ohmeda Medical has 13 total recalls tracked by RecallDetector.

Related Recalls