Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Recalled by Oakworks Inc Due to Weld failure- the bracket attaches the tabletop tilting...

Date: June 29, 2021
Company: Oakworks Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Oakworks Inc directly.

Affected Products

Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219

Quantity: 55 units

Why Was This Recalled?

Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Oakworks Inc

Oakworks Inc has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report