Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Rezum System Recalled by NXTHERA Due to Needle Bond may not be able to withstand...

Date: July 28, 2017
Company: NXTHERA
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NXTHERA directly.

Affected Products

Rezum System, Vapor BPH Ablation System, Model Number D2201-003. NxThera Inc. 7351 Kirkwood Lane Suite 138 Maple Grove, MN 55369 USA. Intended to relieve symptoms, obstructions, and reduce prostate tissue associated with benign prostatic hyperplasia (BPH). It is indicated for men greater or equal to 50 years of age with a prostate volume greater than or equal to 30cm3 and less than or equal to 80cm3. The Rezum System is also indicated for treatment of prostates with hyperplasia of the central zone and/or a median lobe.

Quantity: 113 units total

Why Was This Recalled?

Needle Bond may not be able to withstand the tension force applied when the shaped needle is deployed into and retracted from prostate tissue to deliver treatments.

Where Was This Sold?

This product was distributed to 19 states: AL, AZ, CA, CO, FL, GA, IL, MA, MI, MN, MO, NE, NJ, NY, OH, PA, TN, UT, WA