Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis. Recalled by NxStage Medical, Inc. Due to Ultrafiltration Volume (UFV) may not decrease during treatment-software...

Date: October 29, 2015
Company: NxStage Medical, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NxStage Medical, Inc. directly.

Affected Products

NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis.

Quantity: 2,134 units

Why Was This Recalled?

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Where Was This Sold?

Worldwide Distribution.