Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Low Volume Cartridge IFU (SKU CAR-125-B) used with the following Recalled by NxStage Medical, Inc. Due to There is a potential patient health risk while...

Date: August 27, 2019
Company: NxStage Medical, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NxStage Medical, Inc. directly.

Affected Products

Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1; 2) NxStage VersiHD Cycler, Model Numbers: NX1000-16, and NX1000-16-A; 3) NxStage One, Model Numbers: NX1000-3 and NX1000-3-A; 4) NxStage One, Model Numbers: NX1000-10 and NX1000-10-A;

Quantity: 1803 units

Why Was This Recalled?

There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.

Where Was This Sold?

This product was distributed to 5 states: AL, AZ, CA, NC, VA

Affected (5 states)Not affected

About NxStage Medical, Inc.

NxStage Medical, Inc. has 29 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report