Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Intelli-C: Right side suspended Tabletop Recalled by NRT X-RAY A/S Due to X-ray system C-arm may experience uncontrolled motion if...

Date: April 17, 2024
Company: NRT X-RAY A/S
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NRT X-RAY A/S directly.

Affected Products

Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

Quantity: 35

Why Was This Recalled?

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.

Where Was This Sold?

This product was distributed nationwide across all 50 states.