Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Intelli-C Recalled by NRT X-RAY A/S Due to There is a potential that the drive shaft...

Date: April 27, 2023
Company: NRT X-RAY A/S
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NRT X-RAY A/S directly.

Affected Products

Intelli-C, Right side suspended Tabletop, REF: 03400000

Quantity: 32 units

Why Was This Recalled?

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About NRT X-RAY A/S

NRT X-RAY A/S has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report