Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Recalled by Nova Biomedical Corporation Due to Software- scheduled A-QC analysis could initiate prior to...

Date: June 19, 2017
Company: Nova Biomedical Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nova Biomedical Corporation directly.

Affected Products

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime CCS Analyzer PN 53656 Stat Profile Prime CCS Analyzer with Scanner PN 53657 Stat Profile Prime CCS Comp Analyzer with Scanner PN 52857 Stat Profile Prime CCS Comp Analyzer Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Quantity: 179 units (US) and 904 (OUS) Total 1083

Why Was This Recalled?

Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Nova Biomedical Corporation

Nova Biomedical Corporation has 19 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report