Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Remisol Advance running Software Version 1.12 The Normand Info Remisol Recalled by Normand-Info S.A.S.U. Due to Results from repeated run for WBC (White Bloodcell...

Date: April 30, 2020
Company: Normand-Info S.A.S.U.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Normand-Info S.A.S.U. directly.

Affected Products

Remisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS &). It can also be connected to a host system.

Quantity: 663

Why Was This Recalled?

Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Normand-Info S.A.S.U.

Normand-Info S.A.S.U. has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report