Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Remisol Advance running on hardware with Windows XP Recalled by Normand-Info S.A.S.U. Due to A patch was released for a critical remote...

Date: August 15, 2019
Company: Normand-Info S.A.S.U.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Normand-Info S.A.S.U. directly.

Affected Products

Remisol Advance running on hardware with Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating systems

Quantity: 6479

Why Was This Recalled?

A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Normand-Info S.A.S.U.

Normand-Info S.A.S.U. has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report