Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

nordicICE 2.3.14 Image processing software package used by trained professionals Recalled by NordicNeuroLab AS Due to An error was discovered in the interpretation of...

Date: September 12, 2014
Company: NordicNeuroLab AS
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NordicNeuroLab AS directly.

Affected Products

nordicICE 2.3.14 Image processing software package used by trained professionals, including physicians and medical technicians.

Quantity: 97 licenses

Why Was This Recalled?

An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.

Where Was This Sold?

Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.