Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Temporary Snap Abutment Engaging CC RP 1.5 mm Recalled by Nobel Biocare Usa Llc Due to Packages of abutments may contain an incorrect screw....

Date: August 9, 2019
Company: Nobel Biocare Usa Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nobel Biocare Usa Llc directly.

Affected Products

Temporary Snap Abutment Engaging CC RP 1.5 mm; Pre-manufactured prosthetic components directly connected to the endosseous implant; catalog no. 38761 - Product Usage: Dental implant abutments are intended to be used in the upper or lower jaw and used for supporting tooth replacements to restore chewing function.

Quantity: 478 pieces

Why Was This Recalled?

Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Nobel Biocare Usa Llc

Nobel Biocare Usa Llc has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report