Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NobelParallel Conical Connection NP Recalled by Nobel Biocare Usa Llc Due to The top label of article no. 37965, lot...

Date: August 28, 2019
Company: Nobel Biocare Usa Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nobel Biocare Usa Llc directly.

Affected Products

NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Quantity: 228 pieces

Why Was This Recalled?

The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).

Where Was This Sold?

International distribution - the country Russia only.

About Nobel Biocare Usa Llc

Nobel Biocare Usa Llc has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report