Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Implant Retrieval Instrument CC 3.0 31 mm- Product Usage: They Recalled by Nobel Biocare Usa Llc Due to Due to a problem in the packaging sealing...

Date: March 14, 2019
Company: Nobel Biocare Usa Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nobel Biocare Usa Llc directly.

Affected Products

Implant Retrieval Instrument CC 3.0 31 mm- Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.

Quantity: 31 units

Why Was This Recalled?

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Nobel Biocare Usa Llc

Nobel Biocare Usa Llc has 66 total recalls tracked by RecallDetector.