Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Galaxy System (GAL-001) Recalled by Noah Medical Due to Bronchoscopic system uninterruptable power may have defective component,...

Date: October 25, 2024
Company: Noah Medical
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Noah Medical directly.

Affected Products

Galaxy System (GAL-001)

Quantity: 26

Why Was This Recalled?

Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

Where Was This Sold?

This product was distributed to 16 states: CA, CT, ID, IL, IN, MD, MA, MO, NJ, NY, NC, OH, PA, TN, VA, WV