Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and Recalled by Noah Medical Due to Bronchoscopic visualization system power supplies have defective semiconductor...

Date: October 28, 2024
Company: Noah Medical
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Noah Medical directly.

Affected Products

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Quantity: 18

Why Was This Recalled?

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

Where Was This Sold?

This product was distributed to 15 states: CA, CT, ID, IN, MD, MA, MO, NJ, NY, NC, OH, PA, SC, TN, WV