Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MEE-1000A Neuromaster Recalled by Nihon Kohden America Inc Due to The MEE-1000A Neuromaster had defective printed circuit boards...

Date: January 19, 2012
Company: Nihon Kohden America Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nihon Kohden America Inc directly.

Affected Products

MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography. (EMG)

Quantity: 477 units total (137 units in US)

Why Was This Recalled?

The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Nihon Kohden America Inc

Nihon Kohden America Inc has 10 total recalls tracked by RecallDetector.

Related Recalls