Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NICO Myriad Console Model number NN-7000 The Myriad System is Recalled by Nico Corp. Due to Two cables within the console were inadvertently twisted...

Date: July 29, 2015
Company: Nico Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nico Corp. directly.

Affected Products

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.

Quantity: 6 units

Why Was This Recalled?

Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.

Where Was This Sold?

This product was distributed to 4 states: CA, IN, MD, NC

Affected (4 states)Not affected

About Nico Corp.

Nico Corp. has 7 total recalls tracked by RecallDetector.