Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone Recalled by Nextremity Solutions Due to During insertion of the middle phalanx implant into...

Date: January 13, 2022
Company: Nextremity Solutions
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Nextremity Solutions directly.

Affected Products

Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes SKU: CH-MINI-KT

Quantity: 23 units

Why Was This Recalled?

During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Nextremity Solutions

Nextremity Solutions has 15 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report