Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

StoneLight 30 Laser System (NS3000). For use in surgical procedures Recalled by New Star Lasers, Inc. Due to During internal testing a situation was discovered where...

Date: May 13, 2015
Company: New Star Lasers, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact New Star Lasers, Inc. directly.

Affected Products

StoneLight 30 Laser System (NS3000). For use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.

Quantity: 52 units

Why Was This Recalled?

During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the Graphical User Interface display.

Where Was This Sold?

This product was distributed to 8 states: AL, AZ, CA, FL, IL, MA, MI, OH

Affected (8 states)Not affected

About New Star Lasers, Inc.

New Star Lasers, Inc. has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report