Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Cobra x5 1-Ch EMG ET Tube Recalled by Neurovision Medical Products Inc Due to The electrode insulating material on the Cobra EMG...

Date: July 21, 2020
Company: Neurovision Medical Products Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Neurovision Medical Products Inc directly.

Affected Products

Cobra x5 1-Ch EMG ET Tube, 7.0mm, REF LTE700M-5, Rx Only, Sterile EO CE UDI: B006LTE700M52/$$7061418C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Quantity: N/A

Why Was This Recalled?

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Where Was This Sold?

This product was distributed nationwide across all 50 states.