Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Neurotherm NT2000iX RF Generator Model Number: RFG-NT -2000iX Recalled by Neurotherm, Inc. Due to NT2000iX software shipped with the international setting turned...

Date: October 9, 2013
Company: Neurotherm, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Neurotherm, Inc. directly.

Affected Products

Neurotherm NT2000iX RF Generator Model Number: RFG-NT -2000iX

Quantity: 25

Why Was This Recalled?

NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.

Where Was This Sold?

This product was distributed to 9 states: CA, CO, IA, ME, NJ, NY, SC, TX, WI

Affected (9 states)Not affected

About Neurotherm, Inc.

Neurotherm, Inc. has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report